The Importance of Noise Attenuation Levels in Neonatal Incubators

  1. Fernández-Zacarías, Francisco
  2. Puyana-Romero, Virginia
  3. Hernández-Molina, Ricardo
  1. 1 Universidad de Cádiz
    info

    Universidad de Cádiz

    Cádiz, España

    ROR https://ror.org/04mxxkb11

  2. 2 Acoustic Environments Research Group, Department of Sound and Acoustic Engineering, Universidad de Las Américas, Pichincha, Quito 170125, Ecuador
Revista:
Acoustics

ISSN: 2624-599X

Año de publicación: 2022

Volumen: 4

Número: 4

Páginas: 821-833

Tipo: Artículo

DOI: 10.3390/ACOUSTICS4040049 GOOGLE SCHOLAR lock_openAcceso abierto editor

Otras publicaciones en: Acoustics

Resumen

Background: It is known that high noise levels can be harmful to preterm infants, causing physiological and psychological disorders. It is also known that premature babies spend a lot of time in an incubator. In this sense, many studies show that incubator noise levels can range from 45 to 70 dB. However, these differences in noise levels depend, fundamentally, on the wide range of methodology that can be used. This study aims to know the levels of noise from a fan in the incubator itself and how much it can isolate the noises coming from the outside. Methods: For this purpose, the noise levels of three incubators were measured within a sound-dampened booth for free-field audiometry. For the emission of acoustic energy, a pink noise generator was used; likewise, two microphones were placed, one inside the incubator cabin and the other outside, to determine the acoustic insulation levels of the tested incubators. Results: The incubators produced equivalent continuous sound pressure levels between 53.5 and 58 dB. Acoustic insulation analysis showed that levels varied from one incubator to another, between 5.2 and 10.4 dB. Conclusions: It is necessary to improve the acoustic insulation inside the incubator cabin and to reduce the noise levels of the motor fan. On the other hand, although the incubators are meeting the noise criteria set out in the IEC60601-2-19: 2009 standard of 60 dBA, under normal use conditions, they are still far from the limit recommended by the American Academy of Pediatrics (45 dBA).

Referencias bibliográficas

  • 10.32474/RRHOAJ.2022.07.000267
  • 10.1016/j.earlhumdev.2006.02.009
  • 10.5114/hpc.2022.116199
  • 10.1136/adc.2009.182014
  • 10.1016/0378-3782(77)90035-4
  • 10.3813/AAA.918638
  • 10.4103/1463-1741.199237
  • 10.1186/2052-336X-12-104
  • 10.1111/j.1651-2227.2000.tb00342.x
  • 10.1177/0884217503255009
  • 10.1016/j.pedn.2015.10.011
  • 10.1111/j.1440-1754.2004.00363.x
  • 10.1002/14651858.CD010333.pub3
  • 10.1177/0884217504263293
  • 10.1016/j.jogn.2016.09.005
  • 10.5935/1414-8145.20150014
  • 10.1111/apa.12330
  • 10.1111/jjns.12025
  • 10.1111/jpc.13084
  • 10.1590/S0104-07072012000400007
  • 10.1590/S0104-11692011000500020
  • 10.1891/0730-0832.26.4.213
  • 10.1016/S0755-4982(04)98944-0
  • 10.7196/sajch.676
  • 10.1016/j.apacoust.2020.107301
  • 10.1016/S0147-9563(05)80058-5
  • 10.1590/S0104-11692005000100013
  • 10.1007/s12647-012-0030-0
  • 10.1097/PCC.0b013e31824ea2b7
  • 10.1016/j.pedn.2008.05.002
  • 10.1111/j.1651-2227.2001.tb01355.x
  • 10.1016/S1875-9572(09)60076-0
  • 10.1590/S0103-507X2011000300011
  • 10.1111/jspn.12181
  • 10.1016/j.ijporl.2012.07.001
  • Vendramini, (2011), Acta Paul Enferm., 24, pp. 359
  • 10.3389/fped.2022.955553
  • 10.1016/j.pedn.2007.09.009
  • EN IEC 60601-2-2:2018—Medical Electrical Equipment—Part 2-2: Particular Requirements for the Basic Safety and Essential Performance of High Frequency Surgical Equipment and High Frequency Surgical Accessories https://standards.iteh.ai/catalog/standards/clc/bd56ff83-1df5-41e8-89aa-f89f5abdf25d/en-iec-60601-2-2-2018
  • 4gt Incubadora—Apuntes 1—Guía Tecnológica No. 4 Incubadora Neonatal (GMDN 36025 y 35121) http://www.cenetec.salud.gob.mx/descargas/biomedica/guias_tecnologicas/4gt_incubadora.pdf
  • Berglund, (1999), pp. 1
  • 10.1038/jp.2013.10
  • Etzel, (1997), Pediatrics, 100, pp. 724, 10.1542/peds.100.4.724
  • 10.4103/1463-1741.73999
  • 10.1016/0022-460X(80)90729-4
  • 10.1016/0165-5876(80)90005-1
  • 10.1016/j.ijporl.2018.02.013
  • 10.1016/0022-460X(79)90486-3
  • 10.1590/S0104-11692011000100028
  • 10.1016/j.jcrc.2011.09.003
  • 10.3389/fped.2021.584736
  • 10.1016/j.apacoust.2020.107374
  • 10.3390/children8080704
  • ISO 9612-2009: Acoustics—Determination of Occupational Noise Exposure—Engineering Method https://www.iso.org/cms/render/live/en/sites/isoorg/contents/data/standard/04/17/41718.html
  • 10.3389/fpsyg.2020.01055